SFCN Water Temperature Monitoring Standard Operating Procedure WT01: Testing the Precision of HOBO Loggers in a Cold Bath—Version 1.1

South Florida/Caribbean Network. 2020. SFCN Water Temperature Monitoring Standard Operating Procedure WT01: Testing the Precision of HOBO Loggers in a Cold Bath—Version 1.1. National Park Service

This standard operating procedure (SOP) describes the testing procedures for HOBO® water temperature loggers in the office before field deployment. By definition pure water changes phase from the liquid to the solid state at precisely 0°C, thus one can check the accuracy of a temperature measuring device against a standard consisting of a pure (distilled) water and ice slurry. The objective is to verify that the loggers record the temperature of the slurry to within the ± 0.2°C accuracy range reported for the device (note that this is not a calibration procedure—there is no way to adjust the calibration of the loggers). It is not necessary to perform a warm water test for these devices. Although it is possible to perform such a test, a controlled laboratory setting and high precision instruments would be necessary to provide appropriately accurate and reliable results, and such a test would be time consuming and costly.

Type
Standard Operating Procedure
Authors
South Florida/Caribbean Network
Date of Issue
2020-12
Publisher
National Park Service
Units
BISC , BUIS , DRTO , SARI , SFCN , VIIS
Keywords
Biscayne National Park, Buck Island Reef National Monument, Caribbean, Coral Reef Water Temperature, Dry Tortugas National Park, Logger, National Park Service, Reef Depth, Salt River Bay National Historical Park and Ecological Preserve, South Florida, South Florida/Caribbean Monitoring Network, Standard Operating Procedure, Temperature, U.S. Virgin Islands, Virgin Islands National Park, Water, Water Temperature
Subjects
Ecological Framework: Biological Integrity | Focal Species or Communities | Marine Communities